VS 8/VS 8A Vital Signs Monitor Operator’s Manual B - 1
B EMC and Radio Regulatory Compliance
B.1 EMC
The device meets the requirements of IEC 60601-1-2: 2014.
NOTE
Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this device could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
The non-ME EQUIPMENT (e.g. information technology equipment (ITE)) that
is a part of an ME SYSTEM may be disrupted by the electromagnetic
interference of nearby equipment. It may be necessary to take mitigation
measures, such as re-orienting or relocating the non-ME EQUIPMENT or
shielding the location.
Use of this equipment adjacent to, or stacked with, other device should be
avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to
verify that they are operating normally.
This equipment is intended for use in professional healthcare facility
environment only. If it is used in a special environment, such as magnetic
resonance imaging environment, the equipment/system may be disrupted
by the operation of nearby equipment.
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the this device, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this device
could result.
B - 2 VS 8/VS 8A Vital Signs Monitor Operator’s Manual
If the system is operated within the electromagnetic environment listed in Table
Guidance and Declaration — Electromagnetic Immunity, the system will remain safe
and provide the following essential performance:
Operating mode
Accuracy
Function
Accessories identification
Data stored
Alarm
Detect for connection
Guidance and Declaration – Electromagnetic Emissions
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Emission tests Compliance
Electromagnetic environment
guidance
Conducted and radiated RF
EMISSIONS CISPR 11
Group 1 The device uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby
electronic device.
Conducted and radiated RF
EMISSIONS CISPR 11
Class A The device is suitable for use in all
establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic distortion
EMISSIONS IEC 61000-3-2
Class A
Voltage fluctuations/Flicker
EMISSIONS IEC 61000-3-3
Complies