VS 8/VS 8A Vital Signs Monitor Operator’s Manual 8 - 11
The TemporalScanner™ case can be wiped down using a cloth dampened with
70% isopropyl alcohol. The industrial grade housing and design of the electronic
components allow for completely safe cleaning with 70% isopropyl alcohol but
should not be immersed in fluid or autoclaved.
■ Cleaning the sensor lens
Dirt, greasy films or moisture on the lens will interfere with the passage of infrared
heat and affect the accuracy of the product.
Clean the lens every two weeks with a cotton swab dipped in alcohol. Use only
light force for cleaning, to avoid damaging the lens. Water can be used to remove
any residual film left by the alcohol. Do not use bleach or other cleaning solutions
on the sensor lens.
8.5 Manually Inputting Temp
You can manually input a Temp value and save the data for future reference.
Follow this procedure:
1. Select in the Temp area. For monitors configured with Mindray SmarTemp™
module, to input Temp value in Continuous Monitoring mode, press and hold
in the Temp area.
2. Select a Temp Measurement Site. The options displayed in the menu is
configurable. For details, refer to 16.8.3 Setting Temp Measurement Sites.
3. Input a measured Temp value.
4. Select the Save quick key to save the data.