VS 8/VS 8A Vital Signs Monitor Operator’s Manual 13 - 3
13.6 Clearing Recording Tasks
To clear recording tasks, follow this procedure:
1. Select the Main Menu quick key from the Device column select Record Setup.
2. In the Record Setup menu, select Clear All Record Tasks. This clears all queued
recording tasks and stops the current recording.
13.7 Loading Paper
To load paper, follow this procedure:
1. Use the latch at the upper right of the recorder door to pull the door open.
2. Insert a new roll into the compartment as shown below. Feed the paper through
and pull some paper out from the top of the roller.
3. Close the recorder door.
CAUTION
Use only specified thermal paper. Otherwise, it may cause damage to the
recorder’s printhead, the recorder may be unable to print, or poor print
quality may result.
Never pull the recorder paper with force when a recording is in process.
Otherwise, it may cause damage to the recorder.
Do not leave the recorder door open unless you are reloading paper or
cleaning a paper jam.
13.8 Removing Paper Jam
If the recorder works incorrectly or produces unusual sounds, check if there is a paper
jam. If a paper jam is detected, follow this procedure to remove it:
13 - 4 VS 8/VS 8A Vital Signs Monitor Operator’s Manual
1. Open the recorder door.
2. Take out the paper and tear off the draped part.
3. Reload the paper and close the recorder door.
13.9 List of Recording Reports
The recorder can output the following reports:
Realtime reports
Realtime Segment Waveform Record (Continuous Monitoring)
History reports
Spotcheck Record
EWS Record
Event Detail
Tabular Trend Repor t
Graphic Trend Report