4-42 
5.  Connect the gas cylinder, gas bag and sampling line using a T-shape connector as shown 
below. 
6.  Vent the sampling line to a certain standard gas opening the cylinder pressure relief 
valve. Set the calibration value to the O2 concentration of the standard gas. 
 
7. Select [Calibrate] to start a calibration. 
8.  After a successful calibration, the screen shows [Calibration Completed!]. Otherwise, 
the message [Calibration Failure! Please try again.] is displayed. In this case, you 
need to do the calibration again
.
 
 
4.3.10.3 Commonly-encountered Problems and Recommended Actions   
 
Failure description  Possible cause  Recommended action 
Calibration menu is 
inaccessible. 
The AG module is not fully warmed up 
or not in measure mode. 
Wait for the AG module to be 
fully warmed up and then 
access the calibration menu. 
Calibration is not 
completed. 
1. The module is damaged. 
2. The difference between the set O2 
calibration concentration and the 
selected standard O2 concentration is 
too great. 
1. Return the module to factory 
for repair. 
2. The difference between the 
standard gas concentration and 
the set calibration concentration 
can not exceed 25% of the 
standard gas concentration. 
 
AG module 
Gas bag 
Gas cylinder 
Pressure relie
 valve
Sampling line