6-32
6. Unscrew the four screws fixing the partition plate to remove the partition plate.
Disconnect the connectors from the isolation transformer. Unscrew the four screws to
remove the isolation transformer.
6.2.29 Replace the Caster
Remove the caster and replace with a new one with one person tipping the anesthesia
machine and another person unscrewing the four screws, as shown below.
WARNING
z Replacing a caster requires at least two people. Personal injury and/or machine
damage is possible if one person attempts this procedure alone.