EasyManua.ls Logo

Mindray WATO EX-55 - Check the Flow Sensor Accuracy

Mindray WATO EX-55
322 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-14
4.2.5 Check the Flow Sensor Accuracy
NOTE
z If a great deviation of TV measured value occurs, test the measurement accuracy
of flow sensors so as to determine whether to perform flow calibration again.
To check the measurement accuracy of flow sensors:
1. Make sure that the circuit, calibration device (or other flow measurement device) and
breathing tubes are connected in serial, similar to tubes connection in flow calibration.
For details, refer to4.3.2Flow Calibration (factory).
2. When the system is Standby, select the [Maintenance] shortcut key [Factory
Maintenance >>] [Diagnostic Test >>] [ValvesTest Tool >>] to access the
[ValvesTest Tool] menu.
3. Set PEEP safety valve to [ON].
4. Set the D/A value of the PEEP valve to above 1500 and ensure that the pressure at
which the expiratory valve closes is above 30 cmH2O.
5. Increase the D/A value of the inspiratory valve, causing the measured flow value of the
anesthesia machine calibration device to fall with the following ranges respectively:
(3±0.5) L/min, (10±1) L/min, (20±1) L/min, (30±2) L/min, (60±3) L/min. Record the
measured flow values of the inspiratory flow sensor, expiratory flow sensor and
ventilator flow sensor to which each setting corresponds respectively.
6. Make sure that the deviation between the measured data of the inspiratory flow sensor,
expiratory flow sensor and ventilator flow sensor and that of the anesthesia machine
calibration device must not exceed 1 L/min or 5% of the measured value of the
calibration device, whichever is greater. Otherwise, refer to 4.3.2Flow Calibration
(factory) to perform flow calibration again.
7. If anesthesia machine calibration device is unavailable, you can execute steps 1 through
5 to test the accuracy of flow sensors. The deviation between the measured data of the
inspiratory flow sensor and expiratory flow sensor and that of the ventilator flow sensor
must not exceed 1 L/min or 5% of the measured value of the ventilator flow sensor,
whichever is greater. Otherwise, refer to 4.3.2Flow Calibration (factory) to perform
flow calibration again.

Table of Contents

Related product manuals