5-1 
5 Troubleshooting 
5.1 Introduction 
In this chapter, anesthesia machine problems are listed along with possible causes and 
recommended actions. Refer to the tables below to check the anesthesia machine, isolate 
and eliminate the problems. 
 
Once isolating the part you suspect defective, refer to 6Repair and Disassembly and 7Parts 
to disassemble the equipment and repair and replace the defective part. 
 
5.2 Technical Alarm Check 
Before troubleshooting the anesthesia machine, check for technical alarm message. If an 
alarm message is presented, eliminate the technical alarm first. 
 
The following sections detail how to troubleshoot technical alarms related to the ventilator 
and electronic flowmeter. For detailed information on possible causes and actions for 
technical alarm messages of other modules, refer to the Operator’s Manual. 
 
5.2.1 Ventilator Related Alarms 
In the “Level” column of the following table, the default alarm level is indicated: H for high, 
M for medium, L for low and P for prompt message. 
 
Alarm message  Level  Possible cause  Recommended action 
RT Clock Need 
Reset 
H  There was no button cell 
available in the system, or 
the battery had no capacity.
1. Replace with a new button cell. 
2. If the problem persists, replace 
the main control board. 
RT Clock Not 
Exist 
H  RT chip malfunction.  1. Restart the machine. 
2. If the problem persists, replace 
the board. 
Low Battery 
Voltage! 
H  The battery voltage was too 
low.  
1. The system is operational. 
Connect the AC mains immediately.
2. Check if the battery is normal. 
3. Check if the charging circuit is 
normal. If not, replace the power 
board.