4.2.6 Check the Pressure Sensor Accuracy 
 
NOTE 
z  Generally, measurement deviations do not easily occur to pressure sensors. 
However, in case of maintaining or replacing the monitor board, three-way valve 
assembly, or expiratory valve assembly, you need to perform pressure calibration 
and check the flow sensors accuracy so as to confirm the effectiveness of 
calibration. 
 
To check the measurement accuracy of pressure sensors: 
 
1.  Make sure that the pressure sampling line and calibration device (or other pressure 
measurement device) are connected in parallel, similar to tubes connection in pressure 
calibration. For details, refer to 4.3.3Pressure Calibration (factory). 
2.  When the system is Standby, select the [Maintenance] shortcut key → [Factory 
Maintenance >>] → [Diagnostic Test >>] → [Valves—Test Tool >>] to access the 
[Valves—Test Tool] menu. 
3.  Set PEEP safety valve to [ON]. 
4.  Increase the D/A value of the PEEP valve, causing the measured pressure value of the 
anesthesia machine calibration device to fall with the following ranges respectively: 
(5±1) cmH2O, (20±1) cmH2O, (50±1) cmH2O, (70±2) cmH2O, (90±2) cmH2O. Record 
the measured pressure values of the airway pressure sensor and PEEP pressure sensor to 
which each setting corresponds respectively. 
5.  Make sure that the deviation between the measured data of the airway pressure sensor, 
PEEP pressure sensor and that of the anesthesia machine calibration device must not 
exceed 1 cmH2O or 2% of the measured value of the calibration device, whichever is 
greater. Otherwise, refer to 4.3.3Pressure Calibration (factory) to perform pressure 
calibration again.