6-2
6.1 Prepare for Disassembly
6.1.1 Tools
During parts disassembling and replacing, the following tools may be required:
Metric Allen wrench (2.5#, 3#, 4#, 5#, 8#)
Phillips screwdriver
Diagonal pliers
Flathead screwdriver
Metric M3 and M4 socket screwdriver
Flowmeter calibration fixture
Adjustable wrench
Tweezers
M16 nut mounting fixture (0621-J26-1)
6.1.2 Preparations
Before disassembly,
Make sure that the anesthesia machine is turned off and disconnected from the AC
power source.
Bleed the gas pressure inside the anesthesia machine as described below.
Disconnect all pipeline and cylinder gas supplies.
Prepare the tools required for disassembly.
Maneuver the anesthesia machine to an appropriate location and then step down the
four caster brakes to fix the machine.
CAUTION
z The internal parts may be contaminated during long-term use of the equipment.
Wear special gloves during disassembling and inspecting.