4-62 
 
To adjust the APL valve: 
  (1)Remove the valve cover. 
  (2)Install the locking ring mounting fixture onto the locking ring. 
  (3)Press the drive axis of the new APL valve using a flathead screwdriver and keep 
the screwdriver unmoved. 
  (4)Turn the locking ring mounting fixture counterclockwise to loosen the locking ring. 
  (5)Keep the locking ring mounting fixture unmoved. Turn the flathead screwdriver 
clockwise or counterclockwise to adjust the drive axis of the new APL valve, as shown 
below (turn the drive axis clockwise to increase APL and counterclockwise to decrease). 
  (6)After completing adjustment, keep the flathead screwdriver unmoved. Turn the 
locking ring mounting fixture clockwise to lock the locking ring. 
  (7)Re-install the valve cover. 
 
  
 
 
 
Flathead 
screwdriver 
Locking ring 
mountin
 fixture
APL valve