5-5 
pressure switch and the monitor 
board and check the socket. 
3. If the above two items are 
normal, replace the monitor board. 
O2 Supply 
Failure 
H  The O2 pressure was low.  Use the same method to drive gas 
pressure low to check the O2 
pressure switch. 
Sustained Airway 
Pressure 
H  The airway pressure in the 
breathing circuit was 
greater than sustained 
airway pressure alarm limit 
for 15 seconds.   
Paw<-
10cmH2O 
H  Less than the barometric 
pressure for 10 cmH2O. 
1. Check if the airway pressure 
sensor is in correct measurement 
status.  
2. Check if the sampling line of the 
airway pressure sensor is in normal 
status. 
3. Sensor failure. Replace the 
monitor board. 
ACGO On  M  ACGO was switched on  1. Switch off ACGO. 
2. Check if the connection line of 
ACGO is broken or short circuited. 
3. If the ACGO switch is in correct 
status, and its connection line and 
connector are normal but the 
problem persists, replace the 
monitor board. 
PEEP Valve 
Failure 
M 
Insp Valve 
Failure 
M 
PEEP Safety 
Val v e  F a il u re  
M 
Valve voltage error or 
valve opening status error 
Check if the sensor is normal. 
Replace the sensor or monitor board 
when necessary. 
Replace the expiratory valve 
assembly. 
O2 Flush Failure  M  Oxygen flushing having 
lasted too long (more than 
15 seconds) 
1. Check if the O2 flush button fails 
to pop up. 
2. Check if the O2 flush pressure 
switch is in normal status.   
3. Check if the connection line of 
O2 flush button and its connector 
are normal. 
4. Replace the monitor board. 
Replace O2 
sensor 
M  O2 sensor failure  Replace the O2 sensor.