5-53
5. Check the lifting device and sodalime canister
Tools required:
Anesthesia machine calibration device (quantity: 1)
Injector (quantity: 1)
Φ6 silicone tube (quantity: 2)
PU tube (6X300) (quantity: 1)
Breathing tube (quantity: 3)
Y piece (quantity: 1)
Breathing tube Y piece (quantity: 1)
Breathing tube adapter connector (quantity: 1)
T-shaped Allen wrench (quantity: 1)
Test procedures:
1 Turn off the system switch.
2 Mount the sodalime canister onto the lifting device properly.
3 Remove the lifting device from the patient circuit.
4 Connect the two connectors of the lifting device by using two breathing tubes and
one breathing tube Y piece. The other end of the breathing tube Y piece is connected
to the breathing tube adapter connector through another breathing tube. Connect the
injector connector, pressure sensor (of the anesthesia machine calibration device)
connector (positive pressure end), and the breathing tube adapter connector to a Y
piece, as shown below.