5-65 
4.  If when the D/A value of the inspiratory valve is set to more than “4000”, the flow 
measured by the standard flow measurement device fails to reach 90 L/min, flow 
calibration will be failed. In this case, you need to replace the expiratory valve 
assembly or the monitor board. 
5.  To locate if the DA on the monitor board is faulty, you can use a multimeter to 
measure the output of DA on the monitor board corresponding to the inspiratory valve. 
If voltage also increases with the increase of D/A value, and voltage nears 6V when 
D/A value is set to more than “4000”, it indicates that the DA on the monitor board 
corresponding to the inspiratory valve may be normal. 
6.  After the expiratory valve assembly or monitor board is replaced, you can use the 
similar method to check if the problem is fixed. 
 
5.4.3.5 Problems Related to Opening States of the PEEP Safety Valve 
and PEEP Valve   
When the PEEP safety valve is permanently OFF and the gas supplies are normal, the 
[Drive Gas Pressure Low] is alarmed. When the PEEP valve is faulty, pressure related 
alarms occur in mechanical ventilation state.   
 
By using the valves-test tool, you can detect if the opening states of the PEEP safety valve 
and PEEP valve are normal. 
  To diagnose the opening state of the PEEP safety valve: 
1.  Make sure that gas supplies are normal.   
2. In the [Valves-Test Tool] menu, when the PEEP safety valve is switched on, a subtle 
click is heard. 
3.  Adjust the D/A value of the PEEP valve to cause the pressure measured by the PEEP 
pressure sensor to exceed 0 cmH2O. 
4.  Switch off the PEEP safety valve. The pressure measured by the PEEP pressure sensor 
should drop to 0 cmH2O immediately. Switch on the PEEP safety valve again. The 
measured value of the PEEP pressure sensor rapidly restores almost the same value to 
that before PEEP safety valve is switched off. During this period, gas flow and also 
change of gas flow when the PEEP safety valve is switched on or off can be felt at the 
PEEP outlet, which helps to judge if the PEEP safety valve can be switched on or off 
normally. 
5.  If an error is detected, it is possible that the PEEP safety valve or the safety valve drive 
voltage on the monitor board is faulty. You can use a multimeter to measure the drive 
signals on the monitor board corresponding to the PEEP safety valve (measurement 
can be performed at the corresponding socket). When the PEEP safety valve is turned 
on, the drive voltage should near 6V. When the PEEP safety valve is turned off, the 
drive voltage should near 0V. If these two conditions are met simultaneously, the 
monitor board is normal.