4-23
In the following pictures, A and B are two special communication cables for the calibration
device.
Connection 1: Remove the top cover of the anesthesia machine to expose the monitor
board. Use communication cable A to connect the calibration communication connector
of the calibration device with that of the monitor board, as shown below.
Connection 2: Use communication cable B to connect the calibration communication
connector of the calibration device with that of the anesthesia machine (also the CIS
power connector), as shown below.
Communication cable A
Calibration communication connector
of the monitor board
Calibration communication
connector of the anesthesia machine
Communication cable B