4-28
1. Let the anesthesia machine calibration device be powered. Refer to the method
described in 4.3.2Flow Calibration (factory) to manually zero the calibration device
first. Use the special communication cable to connect the calibration device to the
anesthesia machine.
2. A four-way device is required to connect the sampling lines for pressure calibration. The
following pictures show the four-way device, connectors on the calibration device and
monitor board involved for pressure calibration.
3. Unplug the PEEP pressure sampling line from the PEEP pressure sampling connector on
the monitor board. Then connect it to one connector (Connector A) of the four-way
device.
4. Connect the second connector (Connector B) of the four-way device to the PEEP
pressure sampling connector (high pressure) on the monitor board.
PEEP pressure
sampling line
Airway pressure
sampling line
Four-way device connecting the
sampling lines for pressure calibration
Airway pressure sampling
connector (high pressure)
PEEP pressure sampling connector (high pressure)
FPM-65 pressure sampling connector (high pressure)