4-40 
 
6.  Vent the sampling line to a certain standard gas opening the cylinder pressure relief 
valve. 
7. In the [AG Module Cal.] menu, the measured gas concentration and flow are displayed. 
  If the difference between the measured gas concentration and the actual one is very 
small, a calibration is not needed. 
  If the difference is great, you should perform a calibration. 
8.  Enter the vented gas concentrations.   
9. Select [Calibrate] to start a calibration. 
10.  After a successful calibration, the screen shows [Calibration Completed!]. Otherwise, 
the message [Calibration Failure! Please try again.] is displayed. In this case, you 
need to do the calibration again
.
 
 
NOTE 
z  If the calibration fails, you can select [Defaults] to restore the factory default 
calibration values. If the deviation is great, select [Calibrate] again to do a 
calibration. 
z  If the calibration still fails, replace the AG module. 
 
4.3.9.3 Commonly-encountered Problems and Recommended Actions   
 
Failure description Possible cause  Recommended action 
Calibration menu is 
inaccessible. 
The AG module is not fully warmed up 
or not in measure mode. 
Wait for the AG module to be 
fully warmed up and then access 
the calibration menu. 
Calibration is not 
completed. 
1. The module is damaged. 
2. The difference between the set AG 
calibration concentration and the 
selected standard AG concentration is 
too great. 
1. Return the module to factory 
for repair. 
2. The difference between the 
standard gas concentration and 
the set calibration concentration 
can not exceed 25% of the 
standard gas concentration.