DMC 3000 User's Guide
Major Faults
Display Cause Solution
Pulse counting limitation due to an amount of mecha-
nical shocks or electromagnetic interference on the
DMC3000 (ShKI dEF) or on the module (ShKX dEF).
Conrm whether or not this fault occurs with "normal use"
(no high mechanical or electromagnetic stress).
Contact Mirion for verication.
Electrical problem related to the speaker circuit
(DC 3V or audio frequency).
Verify that the speaker cavity is not obstructed and the
battery voltage is not too low (low battery fault).
Contact Mirion for verication.
Electrical problem or conguration problem of the
DMC3000 and its link to the module.
Put the DMC 3000 in pause then do a "Detach Module"
with DMCUser software, or remove the battery to remove
the faults.
Check that DMC 3000 rmware is more recent than V.7.
Check the connection and connection board then connect
the module.
See §2.2 for compatibility, see §10 for Add-on Modules.
The module is not recognized by the DMC 3000.
Check the conguration of the DMC 3000 that could accept
this type of module.
Minor Faults
Display Cause Solution
Exceeded due date of calibration
Perform a calibration check on DMC 3000 (CALi dAt) or on
the module (CALX dAt).
Missing bit/pointer in the histogram
Verify that the battery voltage is sucient and perform an
entry with DMCUser software.
If the fault persists, contact Mirion for verication.