EasyManua.ls Logo

MOBA CB.D2.CAN - To Transfer Current Sensor Value into Reference;Offset; Zero Reference;Offset Value

MOBA CB.D2.CAN
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© Copyright MOBA Mobile Automation Australia Pty Ltd
Working with Sensors
37
5.3.3 TO TRANSFER CURRENT SENSOR
VALUE INTO REFERENCE/OFFSET
From the working screen, press the [FUNC] key’s Up and Down buttons together to set
reference/offset value to be the same as the current sensor value.
5.3.4 TO ZERO REFERENCE/OFFSET VALUE
Use the [AUX] key to highlight the reference, then press the [ENTER] key, to edit the
Ref.
Press the [AUX] key’s Up and Dn buttons together to zero the Ref value.
The currently selected Ref can also be adjusted from the Standard Screen without
entering Ref menu. Simply press [FUNC] key’s Up or Down button to adjust the value.
Ref2 = +1.25
Ref3 = +2.15
Ref1 = +0.50
Ref2 = +1.25
0.82
Ref = 0.82
Slope [%]Slope [%]
MAN
0.82
Ref = 0.50
Slope [%]Slope [%]
MAN
FUNC
F1
F2
Ref2 = +1.25
Ref3 = +2.15
Ref2 = +1.25
Ref3 = +0.00
AUX
Figure 5.3.3.1
Figure 5.3.4.1

Table of Contents