EasyManua.ls Logo

Molectron EPM2000 - Audible Beeper; Average Batch Size; Backlight

Molectron EPM2000
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming
EPM2000/EPM2000e Reference Guide
32
A
UDIBLE BEEPER
Set or query the status of the Audible Beeper option.
Command Form
SYSTem OPTion AUDio <?|ON|OFF>
Examples
Input Instrument Reply
syst opt aud off none
syst opt aud ? SYSTEM: OPTION: AUDIO OFF
AVERAGE BATCH SIZE
Set or query the Average Mode Batch Size of each channel of the EPM2000.
Command Form
CHannel <A|B|ALL> AVERage BATch_size <?|2..99>
Examples
Input Instrument Reply
ch a aver bat 42 none
ch b aver bat 28 none
ch all aver bat ? CHAN A: AVERAGE: BATCH_SIZE 42
CHAN B: AVERAGE: BATCH_SIZE 28
BACKLIGHT
Set or query the status of the EPM2000's backlight.
Command Form
SYSTem OPTion BACklight <?|ON|OFF>
Examples
Input Instrument Reply
syst opt bac off <
LCD backlight deactivates
>
syst opt bac ? SYSTEM: OPTION: BACKLIGHT OFF