Specifications
EPM2000/EPM2000e Reference Guide
63
ATMOSPHERICS (RELATIVE HUMIDITY)
0 to 95% at or below +30°C (+86°F)
0 to 75% from +31°C to +50°C (+87°F to +122°F)
ATMOSPHERICS (ALTITUDE)
Operating 4570 m (15,000 ft)
Non-operating to 12190 m (40,000 ft)
DYNAMICS (RANDOM VIBRATION)
Operating 0.31 g rms, from 5 to 500 Hz, 10 minutes each axis.
Non-operating 2.46 g rms, from 5 to 500 Hz, 10 minutes each axis.
EMISSIONS
Meets or exceeds the requirements of the following standards:
• EN 50081-1 European Community Requirements
• EN 55011 Radiated Emissions Class A
• EN 50081-1 Conducted Emissions Requirements
Note: To maintain emission requirements when connecting to the serial interface, use
only high-quality, double-shielded (braid and foil) cables. The cable shield must have a
low impedance to the connector housings at both ends.
SUSCEPTIBILITY
Meets or exceeds the requirements of the following standards:
• EN 50082-1 European Community Requirements
• IEC 802-2 Electrostatic Discharge, Performance Criteria B
• IEC 801-3 Radiated Susceptibility 3 V/meter from 27 MHz
to 500 MHz unmodulated.
Performance Criteria: Maximum LCD reading
deviation ±3% of reading.
• IEC 801-4 Electrical Fast Transient/Burst