Programming
EPM2000/EPM2000e Reference Guide
46
TRIGGER LEVEL
Set or query a channel’s trigger level. Values accepted are 2 through 20.
Command Form
CHan <A|B|ALL> TRIGgering LEVel <?|2..20>
Examples
Input Instrument Reply
ch a trig lev 18 none
ch b trig lev 9 none
ch all trig lev ? CHAN A: TRIG: LEVEL 18
CHAN B: TRIG: LEVEL 9
TRIGGER SOURCE
Set or query a channel’s trigger source.
Command Form
CHan <A|B|ALL> TRIGgering SouRCe
<?|INTernal|EXTernalPOSitive|EXTernalNEGative|SLAVe>
Examples
Input Instrument Reply
ch a trig src extneg none
ch b trig src slav none
ch all trig src ? CHAN A: TRIG: SOURCE EXTNEG
CHAN B: TRIG: SOURCE SLAVE
UNITS
Set or query a channel’s measurement units. When setting the EPM2000 into Ratio Mode, the
channel argument indicates the numerator of the fraction.
Command Form
CHan <A|B|ALL> UNITS <?|VOLts|JOUles|WATts|HZ|RATio>
Examples
Input Instrument Reply
ch a units jou none
ch b units vol none
ch all units ? CHAN A: UNITS JOULES
CHAN B: UNITS VOLTS