EasyManuals Logo

Molectron EPM2000 User Manual

Molectron EPM2000
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
Programming
EPM2000/EPM2000e Reference Guide
43
SELF-TEST
Instructs the EPM2000 to perform a self-test and return the result.
Command Form
*TST ?
Example
Input Instrument Reply
*tst ? PASS
SPEEDUP
Set or query the status of the EPM2000 Display Speedup function.
Command Form
SYSTem OPTion SPEEDup <?|ON|OFF|SELective>
Examples
Input Instrument Reply
syst opt speed off none
syst opt speed ? SYSTEM: OPTION: SPEEDUP OFF
STATISTICS BATCH SIZE
Set or query the specified channel’s batch size in Statistics Mode. As with setting the batch size from
the front panel, setting a quantity of 0 results in continuous statistical sampling. Setting this value
automatically resets the EPM2000's statistical batch.
Command Form
CHan <A|B|ALL> STATistics BATch_size <?|0|2..9999>
Examples
Input Instrument Reply
ch a stat bat 6126 none
ch b stat bat 496 none
ch all stat bat ? CHAN A: STAT: BATCH_SIZE 6126
CHAN B: STAT: BATCH_SIZE 496