Programming
EPM2000/EPM2000e Reference Guide
43
SELF-TEST
Instructs the EPM2000 to perform a self-test and return the result.
Command Form
*TST ?
Example
Input Instrument Reply
*tst ? PASS
SPEEDUP
Set or query the status of the EPM2000 Display Speedup function.
Command Form
SYSTem OPTion SPEEDup <?|ON|OFF|SELective>
Examples
Input Instrument Reply
syst opt speed off none
syst opt speed ? SYSTEM: OPTION: SPEEDUP OFF
STATISTICS BATCH SIZE
Set or query the specified channel’s batch size in Statistics Mode. As with setting the batch size from
the front panel, setting a quantity of 0 results in continuous statistical sampling. Setting this value
automatically resets the EPM2000's statistical batch.
Command Form
CHan <A|B|ALL> STATistics BATch_size <?|0|2..9999>
Examples
Input Instrument Reply
ch a stat bat 6126 none
ch b stat bat 496 none
ch all stat bat ? CHAN A: STAT: BATCH_SIZE 6126
CHAN B: STAT: BATCH_SIZE 496