Programming
EPM2000/EPM2000e Reference Guide
41
RANGE, MAXIMUM/MINIMUM
Query a channel's maximum or minimum possible range with the current units, responsivity, head, etc.
Command Form
CHan <A|B|ALL> RANGE <MAX|MIN> ?
Examples
Input Instrument Reply
ch all range max ? CHAN A: RANGE MAX=30J
CHAN B: RANGE MAX=10V
ch all range min ? CHAN A: RANGE MIN=1mJ
CHAN B: RANGE MIN=1mV
RESET
Immediately reset the EPM2000.
Command Form
*RST
Example
Input Instrument Reply
*rst <
unit resets
>
RESPONSIVITY
Set or query the pyroelectric or thermopile responsivity of either channel. Express the value in
decimal or scientific notation.
Command Form
CHan <A|B|ALL> RESPonsivity <THERmopile|PYROelectric> <?|
value
>
Examples
Input Instrument Reply
ch a resp pyro 2200 none
ch b resp pyro 1.331e5 none
ch all resp pyro ? CHAN A: RESP: PYRO 2.200E+03
CHAN B: RESP: PYRO 1.331E+05
ch all resp ther ? CHAN A: RESP: THERMOPILE
4.000E+00
CHAN B: RESP: THERMOPILE
4.000E+00