EasyManuals Logo

Molectron EPM2000 User Manual

Molectron EPM2000
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
Programming
EPM2000/EPM2000e Reference Guide
41
RANGE, MAXIMUM/MINIMUM
Query a channel's maximum or minimum possible range with the current units, responsivity, head, etc.
Command Form
CHan <A|B|ALL> RANGE <MAX|MIN> ?
Examples
Input Instrument Reply
ch all range max ? CHAN A: RANGE MAX=30J
CHAN B: RANGE MAX=10V
ch all range min ? CHAN A: RANGE MIN=1mJ
CHAN B: RANGE MIN=1mV
RESET
Immediately reset the EPM2000.
Command Form
*RST
Example
Input Instrument Reply
*rst <
unit resets
>
RESPONSIVITY
Set or query the pyroelectric or thermopile responsivity of either channel. Express the value in
decimal or scientific notation.
Command Form
CHan <A|B|ALL> RESPonsivity <THERmopile|PYROelectric> <?|
value
>
Examples
Input Instrument Reply
ch a resp pyro 2200 none
ch b resp pyro 1.331e5 none
ch all resp pyro ? CHAN A: RESP: PYRO 2.200E+03
CHAN B: RESP: PYRO 1.331E+05
ch all resp ther ? CHAN A: RESP: THERMOPILE
4.000E+00
CHAN B: RESP: THERMOPILE
4.000E+00