L
Logan JordanJul 31, 2025
Why is my Motic Microscope field of view uneven or only partially visible?
- SStephen VaughanAug 1, 2025
Uneven brightness or a partially visible field of view in your Motic Microscope can occur due to several reasons: The bulb may not be installed properly, so ensure it is correctly installed. The filter slider might be in an intermediate position, so adjust it to the correct position. The phase slider may not be in the click-stop position, so set it correctly. Also, check the condenser mounting and correct it if necessary. Other potential causes include the aperture diaphragm being closed too far (open it), the revolving nosepiece not clicked into position (click it into position), or the optical path selector lever in an intermediate position (adjust it, for trinocular models only).