13 Technical data
Maschinenfabrik Reinhausen GmbH 2023 914001150/12 EN MSENSE
®
DGA 2/3
6
Please contact MR to check the use in case of extreme conditions
7
At the point of measurement
Device properties
Mechanical connection Via ball valve and configurable flange DN50 or DN80
(several flanges available); with safety chain for safe in-
stallation and removal
Measuring pipe length Two different lengths for optimum measuring conditions;
Short measuring pipe: 285mm;
Long measuring pipe: 507mm
Oil extraction connection Luer lock socket on the device;
adapter set for safe and clean oil extraction included in
the scope of delivery
Operation Optional VFD display (suitable for use in daylight) with
3operating keys for displaying measured values and
messages as well as for calling up parameters
Materials used Stainless steel and aluminum;
All external parts and parts in contact with oil are
weather-resistant, resistant to transformer oil and UV-re-
sistant
Gasket material FPM (Viton®)
Available colors (terminal
box)
RAL7033
RAL7038
Vacuum resistance
Pressure resistance
5Pa for 48hours
Max. 400kPa
Degree of protection IP66
Dimensions 435 / 657 x 218 x 264mm (version with short / long
measuring pipe)
Device weight Version with short measuring pipe: approx. 12kg
Version with long measuring pipe: approx. 14kg
(without ball valve / flange / safety chain)
Ball valve weight Approx. 6kg (including flange and safety chain)
Electrical connection
Power supply 95...280VAC, 50/60Hz or 95...280VDC (protected
against polarity reversal)
Power consumption <13W
Overvoltage category III
Connection terminals Voltage supply, relays and analog outputs: 2.5mm
2
,
AWG14
Cable gland 3xM20x1.5 or 3x½“NPT; 8…15mm lead diameter