Page #43 background imageLoading...
Page #43 background image
MSA AUER
MSA
Appendix
Control Unit Models 9010 LCD and 9020 LCD
43
GB
2 M - Manual reset with memorisation of the alarm event
Silencing and reset are performed using one of the Up and/or Down buttons
1 Condition with regular signal
2 Condition with signal past the alarm set-point
3 Same as above, but with acoustic alarm silenced
4 Condition with regular signal waiting for reset
1
3
2
4
LED
Relay/opto-isolator
Horn
: off
: as configured
: deactivated
LED
Relay/opto-isolator
Horn
: flashing 1Hz
: in a state of alarm
: deactivated
LED
Relay/opto-isolator
Horn
: flashing 5Hz
: in a state of alarm
: activated
If the signal returns
past the alarm set-point
LED
Relay/opto-isolator
Horn
: fixed light
: in a state of alarm
: deactivated
if reset with regular signal
If the signal
returns to

Table of Contents