ALTAIR 5X PID
23
Description
US
TWA calculation examples:
1hour exposure of 50ppm:
4hour exposure of 50ppm and 4hour exposure of 100ppm:
12 hour exposure of 100 ppm:
Date Display
Current date appears on the display in the format: MM-DD-YY.
Last cal page
Displays the device last successful calibration date in the format: MM-DD-YY. This page can be de-
activated through MSA Link software or the SETUP - CAL OPTIONS page.
Cal due page
Displays the days until the device's next calibration is due (user selectable). This page can be de-acti-
vated through MSA Link software or the SETUP - CAL OPTIONS page.
Discoverable Mode page
Allows the user to put the device into Bluetooth discoverable mode in order to pair with another device.
This page can be deactivated through the SETUP - INSTRUMENT OPTIONS page.
Change VOC Gas? Page
This page is selectable if the "Menu Enable" feature is ON as described in section 4.5. This page
contains the 10 Favorite PID gases, the All Gases list and the Custom Gas list. An example of this
screen is shown below:
(1 hour x 50 ppm) + (7 hours x 0 ppm)
= 6.25 ppm
8hours
(4 hours x 50 ppm) + (4 hours x 100 ppm)
= 75 ppm
8hours
(12 hours x 100 ppm)
= 150 ppm
8hours
This page can be de-activated through MSA Link software.