ALTAIR 5X PID
33
Operation
US
4.5 Device Setup
The device has provisions to access and modify the following parameters through direct button inter-
face:
- Calibration Options
- Alarm Options
- Instrument Options
These menus can be accessed only from the measure page by pressing and holding the T and
S buttons simultaneously until you are prompted for a password.
The operation is as follows:
(1) Turn the device ON and wait until the measure page appears.
(2) Simultaneously press and hold the T and S buttons for approximately five seconds.
Z The default password is "672".
(3) Enter the first digit by pressing the T or S button and confirm with the q button.
Z The cursor jumps to the second digit.
(4) Enter the second as well as the third digits.
Z Incorrect password: device returns to the Main Page.
Z Correct password: user can enter the Setup mode.
The password can be changed with a PC through the MSA Link software.If the password is forgotten,
it can be reset by using MSA Link software. Contact MSA Customer Service for assistance. The
following Options are available by pressing the T and S buttons:
- Calibration Options - see chapter 4.5
- Alarm Options - see chapter 4.5
- Instrument Options - see chapter 4.5
PASSWORD
000