ALTAIR 5X PID
65
Certification
US
7.2 Marking, Certificates and Approvals According to IECEx
Special Conditions:
The model ALTAIR 5X PID/IR shall be charged by Manufacturer's chargers only (0 - 45 °C) and
opened in a non-hazardous area.
When using the ALTAIR 5X PID/IR in a hazardous area, the device should be be worn or carried on
the body. Do not store the device in a hazardous location. This prevents the possibility of the device
building up an electrostatic charge.
The antenna used for activation of the internal RFID tag with the RF radiation power shall not exceed
6 W for Group I and 2 W for Group IIC.
Capacitance:
Quality Assurance Notification: 0080
Year of Manufacture: see Label
Serial Nr.: see Label
Manufacturer:
Mine Safety Appliances Company, LLC
1000 Cranberry Woods Drive,
Cranberry Township,
PA 16066 USA
Product: ALTAIR 5X PID/IR
IECEx-Type
Examination Certificate:
IECEx FTZU 15.0009X
Type of protection:
IEC 60079-0:2011, IEC 60079-1:2014, IEC 60079-11:2011
IEC 60079-18:2009
Performance none
Marking:
Ex ia I Ma
Ex ia IIC T4 Ga , when combustible XCell sensor is not installed
Ex d ia mb IIC T4 Gb
-20 °C ≤ Ta ≤ +50 °C
Li-Ion Um ≤ 6,7 V DC
IP65
D-Ring: 24pF
Charge contact pins: 17pF
Quality Assurance Notification: 0080
Year of Manufacture: see Label
Serial Nr.: see Label