EasyManua.ls Logo

MSA ALTAIR 5X PID - Certification; ATEX Marking and Approvals

MSA ALTAIR 5X PID
91 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ALTAIR 5X PID
64
Certification
US
7Certification
See device label for the certification that applies to your specific device.
USA and Canada
7.1 Marking, Certificates and Approvals According to the Directive 94/9/EC (ATEX)
Special Conditions:
The model ALTAIR 5X PID/IR shall be charged by Manufacturer's chargers only (0 - 45 °C) and
opened in a non-hazardous area.
When using the ALTAIR 5X PID/IR in a hazardous area, the device should be be worn or carried on
the body. Do not store the device in a hazardous location. This prevents the possibility of the device
building up an electrostatic charge.
The antenna used for activation of the internal RFID tag with the RF radiation power shall not exceed
6 W for Group I and 2 W for Group IIC.
Capacitance:
USA
USA / NRTL
(Intrinsic Safety,
Non-Mining)
UL913 for Class I, Div. 1, Groups A, B, C and D
-20 °C to +50 °C, T4
Canada
Canada / CSA
(Intrinsic Safety,
Combustible
Performance, Non-
Mining)
CSA C22.2 No. 157 for Class I, Div. 1, Groups A, B, C and D
CSA C22.2 No. 152 M1984 Combustible Performance
Tamb = -20 °C to +50 °C, T4 for Intrinsic Safety
Tamb = -20 °C to +50 °C, for Combustible Performance
Manufacturer:
Mine Safety Appliances Company, LLC
1000 Cranberry Woods Drive,
Cranberry Township,
PA 16066 USA
Product: ALTAIR 5X PID/IR
EC-Type Examination Certificate: FTZU 15 ATEX 0038X
Standards:
EN 60079-0:2012, EN 60079-1:2014, EN 60079-11:2012
EN 60079-18:2009, EN 50303:2000
Marking:
I M1 Ex ia I Ma
I M1 Ex db ia I Ma - when MSH2ia sensor is installed
II 2G Ex db ia mb IIC T4 Gb
II 1G Ex ia IIC T4 Ga - when MSH2ia and XCell Ex sensor
is not installed
-20 °C ≤ Ta ≤ +50 °C
Li-Ion Um 6,7 V DC
IP 65
D-Ring: 24pF
Charge contact pins: 17pF

Table of Contents