EasyManua.ls Logo

MSA ALTAIR 5X PID - Page 45

MSA ALTAIR 5X PID
91 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ALTAIR 5X PID
45
Operation
US
Zero Calibration
(1) Press the S button for five seconds in Normal Measurement page.
Z ZERO screen displays.
To skip the ZERO procedure and move directly to the span calibration proce-
dure, push the S button. If no button is pushed for 30 seconds, the device
prompts user to perform a SPAN calibration before device returns to the
Normal Measurement page.
To perform ONLY a Fresh Air Setup at this time, press the q button. The
device then performs a Fresh Air Setup as described in chapter 4.2. When
the Fresh Air Setup is complete, the device returns to the normal Measure
screen.
(2) Press the T button to confirm the ZERO screen, i. e. to execute zero
calibration.
Z The message "SENSOR REFRESH" displays, followed by the
message "ZERO CALIBRATION".
Z The "REFRESH" message does not appear if a catalytic combus-
tible sensor is not installed.
Z ZERO calibration starts.
Z A progress bar shows the user how much of the calibration has
been completed.
During the first moments of a ZERO calibration, the combustible sensor
reading may be replaced by a moving display of “PLEASE WAIT”. This is
normal.
Z After the ZERO calibration is completed the device displays either
ZERO CALIBRATION PASS
or
ZERO CALIBRATION FAIL”.
Z Only if the device passes the zero calibration the SPAN screen
displays.

Table of Contents

Related product manuals