EasyManua.ls Logo

MSA SOLARIS - Sensor Missing Alarm

MSA SOLARIS
162 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Failure to follow this procedure, could result in serious per-
sonal injury or death.
During "Battery Low" condition, prepare to exit the work
area since the instrument could go into "Battery Shutdown"
at any time, resulting in loss of sensor function. Depending
on the age of the batteries, ambient temperature and other
conditions, the instrument "Battery Low" and "Battery
Shutdown" times could be shorter than anticipated.
Recharge the instrument or replace the batteries when the
"Battery Low" or "Battery Shutdown" conditions occur.
Sensor Missing Alarm
The Solaris Multigas Detector will enter the Sensor Missing alarm if the
instrument detects that an enabled sensor is not properly installed in the
instrument. For O
2
, CO, and H
2
S sensors, the Sensor Missing feature
is checked when the instrument is turned ON and when leaving the
Setup mode. The combustible Sensor Missing feature is continually
monitored. If a sensor is detected as missing, the following occurs:
SENSOR and MISSING flash on the display
The flag above the sensor detected as missing flashes on the
display
Alarm sounds and lights flash
Alarm can be silenced by pressing the RESET button
No other pages can be viewed
After approximately one minute, the instrument automatically turns
OFF.
If a Sensor Missing condition occurs, stop using the instru-
ment; it can no longer alert you of potential hazards.
1. Leave the area immediately.
2. Turn OFF the instrument if it is ON.
3. Report to the person responsible for maintenance
Failure to follow this procedure could result in serious per-
sonal injury or death.
""
WARNING
""
WARNING
""
CAUTION
2-5

Table of Contents

Related product manuals