push the RESET button.If no button is pushed for 30 seconds,
the instrument returns to the Measure mode.
8. Remove the calibration cap.
9. Close the valve on the regulator.
NOTE: The autocalibration procedure adjusts the span value for any
sensor that passes the test; sensors that fail autocalibration are
left unchanged.
Since residual gas may be present, the instrument may briefly
go into an exposure alarm after the calibration sequence is
completed.
Autocalibration Failure
If the Solaris Multigas Detector cannot calibrate one or more sensor(s),
the instrument goes into the Autocalibration Failure Page and remains
in alarm until the RESET button is pushed. Sensors that could not be
calibrated are indicated by dashed lines on the concentration display.
4-5