Table 6-9. HYDROGEN SULFIDE (appropriate models only)
- Typical Performance Specifications
RANGE 200 ppm H
2
S
RESOLUTION 1 ppm H
2
S, for 3 to 200 ppm H
2
S
REPRODUCIBILITY +2 ppm H
2
S or 10% of reading, whichever is greater
0 to 100 ppm H
2
S, +15% > 100 ppm H
2
S
(normal temperature range*)
+5 ppm H
2
S or 20% of reading, whichever is greater
(extended temperature range*)
RESPONSE TIME 90% of final reading in 60 seconds* (normal temperature range)
*See TABLE 6-2 NOTE
Table 6-10. HYDROGEN SULFIDE - Cross Reference Factors
for Solaris Calibration Using Calibration Cylinder (P/N10045035)
NOTE: Data is presented as the indicated output in ppm, which would result from the
application of 100 ppm of the test gas
TEST GAS (100 PPM) EQUIVALENT PPM
Hydrogen Sulfide (H
2
S) 100 +10
Ethylene (C
2
H
4
)0 +0
Methane (CH
4
)0 +0
Hydrogen (H
2
)0 +0
TEST GAS (100 PPM) EQUIVALENT PPM
Ammonia (NH
3
)0 +0
Chlorine (Cl
2
)0 +0
Nitrogen Dioxide (NO
2
) -20 +2
Nitric Oxide (NO) 1 +1
Carbon Monoxide (CO) 4 +4
Hydrogen Chloride (HCl) 0 +0
Hydrogen Cyanide (HCN) 1 +1
Sulfur Dioxide (SO
2
) 10 +3
Ethanol (EtOH) 0 +
0
Toluene 0 +0
6-7