12 MSI-7300 Operator’s Manual
2.0 Operation
2.1 Power
To turn on the Power
1. Press
POWER
.
The LCD will show all segments for a display test.
The software version number will display.
The D
yna-Link 2 is ready for use.
2.2 Zero
Takes out small deviations in zero when the Dyna-Link 2 is unloaded. See Section 3.6
for zeroing (Taring) package or pallet weights.
Note
Note
The tension reading must be stable within the motion window for the
zero function to work.
1. Press
ZERO
. The display reads 0 (or 0.0 or 0.00, etc).
The backup memory stores the zero reading, and can restore it even if
power fails.
Zero - Rules for Use:
Works in GROSS mode or NET mode. Zeroing while in NET mode will zero
the gross tension causing the display to show the negative tare value.
The unit must be stable within the motion window, it will not zero until the
stable ann
unciator is on.
The unit will “remember” that it has a zero
request for two seconds. If motion
clears in that time, it will zero.
The unit will accept a zero setting over the full range of the Dyna-Link. Zero
setti
ngs above 4% of full Dyna-Link will subtract from the overall capacity.
For example, if you zero out 100 lb on a 1000 lb Dyna-Link, the overall
capacity will reduce to 900 lb plus the allowed over-range amount.