Chapter 2 Analog Input and Temperature Sensor Measurements
Using Differential Connections for Ground-Referenced Signal
Sources
Figure 2-10 shows how to connect a ground-referenced signal source to the DAQ device
configured in differential mode.
Figure 2-10. Differential Connections for Ground-Referenced Signal Sources
With this type of connection, the NI-PGIA rejects both the common-mode noise in the signal and
the ground potential difference between the signal source and the device ground, shown as V
cm
in the figure.
AI+ and AI– must both remain within ±11 V of AI GND.