EasyManuals Logo
Home>National Instruments>Measuring Instruments>USB-621 Series

National Instruments USB-621 Series User Manual

National Instruments USB-621 Series
205 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
Chapter 8 Counters
NI USB-621x User Manual 8-8 ni.com
Note that if you are using an external signal as the Source, at least one
Source pulse should occur between each active edge of the Gate signal.
This condition ensures that correct values are returned by the counter. If this
condition is not met, the counter returns a zero. Refer to the Duplicate
Count Prevention section for more information.
For information about connecting counter signals, refer to the Default
Counter/Timer Pinouts section.
Semi-Period Measurement
In semi-period measurements, the counter measures a semi-period on its
Gate input signal after the counter is armed. A semi-period is the time
between any two consecutive edges on the Gate input.
You can route an internal or external periodic clock signal (with a known
period) to the Source input of the counter. The counter counts the number
of rising (or falling) edges occurring on the Source input between
two edges of the Gate signal.
You can calculate the semi-period of the Gate input by multiplying the
period of the Source signal by the number of edges returned by the counter.
Single Semi-Period Measurement
Single semi-period measurement is equivalent to single pulse-width
measurement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments USB-621 Series and is the answer not in the manual?

National Instruments USB-621 Series Specifications

General IconGeneral
BrandNational Instruments
ModelUSB-621 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals