Chapter 8 Counters
NI USB-621x User Manual 8-30 ni.com
• AI Reference Trigger (ai/ReferenceTrigger)
• AI Start Trigger (ai/StartTrigger)
In addition, Counter 1 Internal Output, Counter 1 Gate, Counter 1 Source,
or Counter 0 Gate can be routed to Counter 0 Aux. Counter 0 Internal
Output, Counter 0 Gate, Counter 0 Source, or Counter 1 Gate can be routed
to Counter 1 Aux.
Some of these options may not be available in some driver software.
Counter n A, Counter n B, and Counter n Z Signals
Counter n B can control the direction of counting in edge counting
applications. Use the A, B, and Z inputs to each counter when measuring
quadrature encoders or measuring two pulse encoders.
Routing Signals to A, B, and Z Counter Inputs
Each counter has independent input selectors for each of the A, B, and Z
inputs. You can route any input PFI signal to each input.
Counter n Up_Down Signal
Counter n Up_Down is another name for the Counter n B signal.
Counter n HW Arm Signal
The Counter n HW Arm signal enables a counter to begin an input or output
function.
To begin any counter input or output function, you must first enable, or
arm, the counter. In some applications, such as buffered semi-period
measurement, the counter begins counting when it is armed. In other
applications, such as single pulse-width measurement, the counter begins
waiting for the Gate signal when it is armed. Counter output operations can
use the arm signal in addition to a start trigger.
Software can arm a counter or configure counters to be armed on a
hardware signal. Software calls this hardware signal the Arm Start Trigger.
Internally, software routes the Arm Start Trigger to the Counter n HW Arm
input of the counter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments USB-621 Series and is the answer not in the manual?

National Instruments USB-621 Series Specifications

General IconGeneral
BrandNational Instruments
ModelUSB-621 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals