Chapter 3 Individual Operations
165
Individual
Operations
(7) Maintenance
(7-1) FL Turret Drive Speed
(See DS-L3 Chapter 6 “6.5 Setting the Driving Speed of the Epi-Fluorescence Cube Turret”.)
Set the drive speed of the epi-fluorescence cube turret to High or Low.
When a thick dichroic mirror is used for a filter cube, set the drive speed to “Low”.
If you are using a two-layered epi-fluorescence cube turret, configure 1st and 2nd layers separately.
In order to perform this configuration, the motorized epi-fluorescence cube turret must be attached.
<Factory setting>
High for both the 1st and 2nd layers
(7-2) Resetting Data
(See DS-L3 Chapter 6 “6.6 Restoring the Factory Default Settings”.)
Restore all settings made on the [MICROSCOPE SETUP MENU] to the default settings.
(7-3) Program Version
You can check the program version of the microscope system.
When you are using an ergo controller, the program version of
the ergo controller is displayed after [ERGC:].
When you are using the joystick controller, the program version
of the joystick controller is displayed after [JOY:].
[SETUP MENU] → [MAINTENANCE]
Maintenance Item Setting
Restoring the Factory Default Settings for Microscope Data (from the Microscope)
You can restore the setting data to the default setting with the button operation of the Ni-E microscope.
Use this method to restore objective device information to factory default.
<Procedure>
Turn on the power while holding down the following three
buttons on the front of the main body.
DISPLAY Previous button
DISPLAY Next button
Z-RESET button
Turning power switch ON while holding down three
buttons
When successful, the display shown on right side appears
and data is restored to factory default.
FLASH Data Initializing.
Setting Data Restored to Default Setting
Power switch ON
(7-3)
(7-1)
(7-2)