REASSEMBL Y PROCEDURE
e~~~
'123f109
f
105
i
106)
. '--.__/
"-._/'
\.___./
1
; I \
Fig.
37
R-4104.A
1.
Coaxial
Coarse
and
Fine Focus Knob
Unit
7) Attach
gears
e@
and @
to
pinion
case
@ ,
and
coarse
focus
knob bearing @
(this should
be
cleaned, beforehand)
to
e@
by
means
of
screws
@x3.
e@
and
e@
are
to
be
oiled
as
shown
in
Fig.
36, after cleaned
and
dried.)
8) lnsert the subassembly consisting of
e@])
(coarse
focus
case)
and
@ (coarse
focus
shaft) and @ and @ (gears),
as
shown
in
Fig.
37
(apply
oil
to
@
as
shown
in
Fig.
10).
9) Thereafter, following
4)-
l
),
reverse
the disassembly procedure. (After cleaned and dried,
fine
focus
shaft @ ,
gears
@ and
@,
@,
e@
and @
and
gears
@
and @ should
be
oiled
as
shown
in
Fig.
24.
Cover
plate @
is
to
be
cemented
by
No.
616.
Apply
G7920
to
plate spring e@ and
washer
@.
-
33
-