EasyManua.ls Logo

Nikon Labophot - Page 94

Nikon Labophot
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
·""-·,
R-4106.A
REASSEMBL
Y
PROCEDURE
1.
B·tube
REASSEMBLY
PROCEDURE
Before reassembly, clear
off
dírt from the lenses, prisms and metal parts.
1.
B-tube
1-1.
Depression prism unit (Refer to Fig.
1)
1) Insert the lens chamber
G)
(with @ and @ ins1de, to be positioned
as
~~
thicker),
and screw in retaining ring
G)
Applying adhesive No. 350, attach
G)
to
G)
aft;;';djusting
the
optical
axis (P. 40).
2) Assemble prism @ on the
mount
(D and
n.x
it with plastic
screw@
transitorily. Set the
prism in such a position
as
shown in Fig. 16,
that the end surface
of
the prism holder coin·
cides with the
bottom
edge
of
the 30° surface
of
the prism
@.
(Fig. 16)
3) Attach
G)
to main body
G)
with screws
~x3
' Coincide the edgcs.
Fig.
16
4) Insert lemes ª and @
into
main body
(D
and screw
in
retaining
ringG)
(the positions
of
the lenses are to be
~J..
thicker).
~
1-2.
Binocular
body
(Refer to Fig. 2 and 3)
1)
Insert
washer@
into the central axle
(@J).
(Apply oil G7920 to
both
the surfaces).
2) Insert righthand prism
box
c:@D
into the central
a.xle
e@
(Apply oil G7920)
to
the fit and
thrwt
surfaces).
Note: Do
not
apply so much oil
as
to
enter
the prism box. Over-full oil should be wiped off.
3)
Insert rotation limit screw
~
x 1 into the groove on e@ from the outside
of@.
Check
the
full range
of
operation
from limit
to
limit.
4)
Insert
washer@
into the central axle
(@D.
(Apply oil G7920
to
both
the surfaces).
-
24-

Table of Contents

Other manuals for Nikon Labophot

Related product manuals