Fehler! Verwenden Sie die Registerkarte 'Start', um Title;Title dem Text zuzuweisen, der hier angezeigt werden soll. |
12 | 04.10.2021 | Page 41
11.
Let the system run until no more disinfectant can be
detected at the permeate drain. Use Minncare residual
test strips for this.
12.
Close the shut-off valve permeate to the drain.
13.
Check that all dialysis stations are free from disinfectants.
Switch the system off during disinfection with a conductivity alarm,
clear the alarm and restart the system.
RISK OF DEATH !
Before the next dialysis, each sampling point must be tested again to ensure
that it is free from
disinfectants.
* Free of disinfection means 0ppm- no discoloration of the test strip.
See the colour scale on the packaging of the test strip Residual.