EasyManua.ls Logo

Oakton WP PCD 650 - System Setup

Oakton WP PCD 650
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instruction Manual PCD 650
67
3.10.2 To calibrate 0% saturation
Figure 60 : 0% saturation calibration Screen
1. Press MODE key to select % Saturation mode.
2. Immerse the DO probe in 0% solution. Stir gently to create a homogenous
solution.
3. Press CAL key to calibrate the meter. The meter shows ‘Dissolved O
2
Calibration-Rinse Electrode’ screen [Figure 48(a)] for few seconds to
prompt user to rinse electrode with de-ionized water before calibration.
4. The meter then shows saturation calibration screen [Figure 60].
5. Wait for the reading to stabilize.
6. The primary display will show the current value of measurement.
7. Press ENTER key to confirm the calibration. The meter automatically
calibrates to 0% saturation and returns to the measurement mode [Figure
59].
Note: The keys that appear in 0% saturation calibration mode functions similar
to the one as described in 100% saturation mode. (Refer page 66)
3.10.3 % DO Calibration Report
Calibration report gives you detailed information for the % saturation. It
includes date & time, 0% and 100% saturation mV value, % saturation offset
calibration and barometric pressure at which the solution was measured.
Note: If you wish to completely re-calibrate the meter, you need to clear
previous calibration data. Press CLR-C (F2) key to clear previous
calibration. The meter shows you confirmation screen. Press ENTER key
to confirm deleting previous calibration. It will clear all the calibration
data including mg/L data.

Table of Contents

Related product manuals