6.4 Installation
For operation the Easygraph must be fastened with
the enclosed holder on a slitlamp.
This holder is designed for slitlamps with a drilling of
8mm diameter. It is possible to get other holders on
request.
For this purpose first screw the enclosed holder with
the knurled screw on the mounting thread (Item 9) of
the Easygraph.
Remove a cover of the support possibly present on
the slitlamp. Then set the shaft of the Easygraph
holder in the receptacle of the slitlamp.
As an alternative the Easygraph can also be mounted
on an optionally available guide rail.
The Easygraph can be pushed with this into a rest
position, so that examinations can then be made with
the slitlamp. However, note that the guide rail can
also move when moving the slitlamp and the
microscope attachment. Hold the guide rail or the
Easygraph firmly if necessary so that they do not
collide with the head support.
6.5 Electrical Connection
The Easygraph may be operated only with the
supplied plug-in power supply unit.
To adapt to different countries the plug-in power
supply unit has replaceable plug adapters for
connection to the mains and a wide range input.
Select the adapter corresponding to your country
and push it into the plug-in power supply unit until it
snaps into place.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the OCULUS Easygraph 70620 and is the answer not in the manual?

OCULUS Easygraph 70620 Specifications

General IconGeneral
BrandOCULUS
ModelEasygraph 70620
CategoryMedical Equipment
LanguageEnglish

Related product manuals