CD Indicators
4.1.4 Calibration Test
Calibration test feature allows a check of a known calibration mass against the last stored calibration information in the
indicator. Sample display illustrates a 40 kg scale base using a 10kg mass.
Procedure
Press the Setup button, SCALE is displayed.
Press button UNTIL CAL is displayed.
Press Enter button, CAL TYPE is displayed.
Press Enter button, SPAN is displayed.
Press button to select calibration test,CAL
TYPE CALtESt is displayed.
Press Enter button, TEST WT is displayed and the
last calibration mass value which was entered with the
first digit flashing. (Sample illustrates 10kg).
Press and or and enter the
desired mass value when using Model CD-31. The
numerical keypad can be used for Model CD-33. This
number must be at least 10% of the full span value.
Press Enter button, CLR PAN is displayed.
Press Enter button,WORKING is momentarily dis-
played, then PUT WT 10 kg is displayed.
Place specified calibration mass on pan.
Press Enter button, WORKING is displayed. After a
few seconds, WEIGH DIFF is displayed.The display
now indicates the actual difference in weight between
the previous weight value which was stored in the
Indicator. After approximately 8 seconds, the display
returns to the WEIGH mode and indicates the weight
on the pan.
• Remove calibration mass from the pan.