EN-48 PX Series Balance
9.4 Communication
9.4.1 Interface Commands
Commands listed in the following table will be acknowledged by the balance.
Command
Characters
Function
IP
Immediate Print of displayed weight (stable or unstable).
P
Print displayed weight (stable or unstable).
CP
Continuous Print.
SP
Print on Stability.
H
Enter Print Header Lines
Z
Same as pressing Zero Key
T
Same as pressing Tare Key.
xT***
Establish a preset Tare value in displayed unit. X = preset tare value. Sending 0T clears tare
(if allowed).
PT
Prints Tare weight stored in memory.
ON
Brings out of Standby
OFF
Goes to Standby.
C
Begin Span Calibration
IC
Begin Internal Calibration, same as trigger from calibration menu.
AC
Abort Calibration. Attention: when LFT ON, the operation is not allowed.
PSN
Print Serial Number.
PV
Print terminal software version, base software version and LFT ON (if LFT is set ON).
x#
Set Counting APW (x) in grams. (must have APW stored)
P#
Print Counting application APW.
x%
Set Percent application reference weight (x) in grams. (must have reference weight stored)
P%
Print Percent application reference weight.
xRL
0 = disable response; 1 = enable response. This command only controls the “OK!” response.
xT
Pre-tare the container weight (x) in grams.