5. Hardware troubles records
This displays, for each material incident detected: incident ID, status
(activated = ON or deactivated = OFF) as well as the date and time of
occurrence or release of the event.
Previous page, Next page, and Last page allow access to the corresponding
pages of the file: each page can display a maximum of 8 lines.
Digital module no longer responding (line cut, module failure, wrong address,
module absent).
Configuration or module address error.
Internal temperature of the MX 43 higher than maximum tolerated value.
Internal temperature of the MX 43 lower than maximum tolerated value.
Switching to external power supply.
Incident on line 1 (short-circuit).
Incident on line 2 (short-circuit).
Incident on line 3 (short-circuit)
Incident on line 4 (short-circuit).
Incident on line 5 (short-circuit).
Incident on line 6 (short-circuit).
Incident on line 7 (short-circuit).
Incident on line 8 (short-circuit).
Calibration defect (zero shifted).
Calibration defect (used cell).
Calibration defect (cell oversensitive).
Calibration defect (measurement unstable).
Table 16: Material incidents file messages.
6. System troubles records
This displays the events relative to MX 43 operation (power failure/fluctuation,
On/Off, etc.).
Previous page, Next page, and Last page allow access to the corresponding
pages of the file; each page can display a maximum of 8 lines.
Failure of internal tests
Contact Post-Sales Service
Table 17: System incidents file messages.
3. Slave info
These data enable maintenance technicians to visualize the communication
framework between MX 43 and the digital modules.