Test Programs Maintenance Manual
3-8
Page
Test No./
Test Name
Function Outline
Press the [Start] button to pause the operation at the break point.
The speed can be changed during the pause.
When an error occurs, the loader displays the same error code as in the regular operation, and then
stops.
Press the [Quit] button, and the A-arm returns to its initial position and the loader stops the operation.
No. 3
A-arm
horizontal
movement
check
<A-arm horizontal movement check>
This program can be used to check the A-arm horizontal movement.
Select the operation speed using the menu buttons on the liquid crystal panel.
[M1]: High speed <Fast speed> [M2]: Medium speed <Middle speed> [M3]: Low speed <Slow
speed>
If you do not specify anything here, the loader will operate at the high speed [M1]: High speed
<Fast speed> setting.
Press the [Start] button.
If the loader has not been initialized, initialization starts.
Press the interlock switch and the [Start] button.
FOUP Door open. (AL120-LMB12-LP3 only.)
When the door opens completely, release your hand from the interlock switch, and press the [Start]
button. ( to are for AL120-LMB12-LP3 only.
Moves to upper limit height.
Center of Macro table Cassette side
Cassette side Center of Macro table
Center of Macro table Standby position
Standby position Cassette side
Cassette side Standby position
Standby position Center of Macro table
The loader then repeats steps to .
The operation will not start if a cassette is set on the elevator.
Press the [Start] button to pause the operation at the break point.
The speed can be changed during the pause.
When an error occurs, the loader displays the same error code as in the regular operation, and then
stops.
Press the [Quit] button, the A-arm returns to its initial position. Then, if you press the [Start] button or
the [Quit] button while pressing the interlock switch of the FOUP opener, the FOUP door closes and
the loader stops the operation.
A-arm moves horizontally to the cassette installation part at high
speed.
When the FOUP Door opening and shutting part is carelessly looked
into, it is likely to injure in A-arm that moves.
Do not continue the operation for longer than 5 minutes.
Wait for longer than one minute before restarting the operation.