Test Programs Maintenance Manual
3-26
Page
No. 22
Button
operation
check
<Button motion check>
This program can be used to check the operation button motions.
Press the [Start] button.
Press a button with an LED, and the LED lights up on the pressed button.
Press the All/Sampling button, and all the LEDs light up.
Press the menu buttons [M1] to [M4] on the liquid crystal panel, and [M1] to [M4] appear on the
liquid crystal panel.
When the 4-way button on the liquid crystal panel is pressed, [UP], [DOWN], [LEFT] or [RIGHT]
corresponding to the button pressed is displayed on the liquid crystal panel.
When the [wafer alignment ] control is manipulated, [Top], [Right], [Bottom] or [Left] corresponding
to the set position is displayed on the liquid crystal panel.
When the [Inspection time] control is manipulated, the inspection time corresponding to the set
inspection time is displayed on the liquid crystal panel.
When the [Top macro spin direction control] is manipulated, [CCW], [Cancel] or [CW] corresponding
to the set rotation is displayed on the liquid crystal panel.
When the remote control buttons are manipulated, [Remote start], [Remote Unload], [Remote
Micro] or [Remote Macro] corresponding to the settings is displayed on the liquid crystal panel.
Press the [Quit] button, and the loader returns to its initial state and stops the operation.
No. 23
Sensor
and
button
check
<Sensor and button input check>
This program can be used to check the sensor input states.
Press the [Start] button.
The LEDs on the Wafer No. selector buttons light up in accordance with the sensor input states.
The list of sensors is displayed on the liquid crystal panel.
Use the menu buttons [M1], [M2] [M3]and [M4] to switch the sensors to check.
Press the [Quit] button, and the loader returns to its initial state and stops the operation.
M1
M2
M3
M4
1
A front/back cassette position sensor
<AL cst>
U205
AL120-LMB12-F lower-limit sensor
<E low> U104
E step-out sensor
<E out2>
U105
2
A front/back standby position sensor
<AL mid>
U206
Horizontal origin sensor
<E limit> U101
E step-out sensor
<E out1>
U106
3
A front/back macro position sensor
<AL macro> U207
Horizontal mapping position sensor
<E map> U102
A step-out sensor
<A out2>
U019
4
A vertical upper-limit sensor
<A up>
U201
Horizontal stop position sensor<E
out> U103
A step-out sensor #2
<A out1>
U019
5
A vertical transfer position sensor
<A mid>
U202
L-arm rotation origin sensor
L rotate> S101
-
6
A vertical lower-limit sensor
<A low> U203
L vertical upper-limit sensor
<L up> S101
7
A side-to-side origin sensor
<A LR org> U208
L vertical transfer position sensor
<L mid>
8
L vertical lower-limit sensor
<L low>
Macro table rotation motor driver
MO <M MO>
9
Cassette switch #
<Caseet>
S101
Macro table rotation motor driver
TSD <M TSD>
10
Cassette switch #
<Caseet> S102
A-arm rotation motor driver MO
<AR MO>
11
Cassette switch #
<Caseet> S103
A-arm rotation motor driver TSD
<AR TSD>
12
Cassette switch #
<Caseet>
S104
A-arm horizontal motor driver MO
<AL MO>
13
A-arm horizontal motor driver TSD
<AL TSD>
14
Window open/close sensor
<W close>U701
A-arm vertical motor driver MO
<AUP MO>
15
Window lock
<WL close>
U702
A-arm vertical motor driver TSD
<AUP TSD>
16
Window unlock
<WL open> U703
17
F horizontal macro position sensor
<FL macro> U505
L reverse rotation switch
<L SW BK>
18
F horizontal middle position sensor
<FL mid>U506
L rotation switch
<L SW FOR>
19
F horizontal stage position sensor
<FL stage> U507
L position registration switch
<L set1>
20
F vertical upper-limit sensor
<F up> U501
21
F vertical middle point sensor
<F mid>U502
Macro table upper-limit sensor
<M up>
L-arm rotation motor driver MO
<L MO>
22
F vertical lower-limit sensor
<F low> U503
Macro table lower-limit sensor
<M low>
L-arm rotation motor driver TSD
<L TSD>
23
<FUP vcm>
Wafer sensor vertical motor driver
TSD <E TSD>