Phenomena Cause Remedy
Reference
page
Wrong cable is used. Use a correct cable. 23
Wrong command is used. Refer to the command reference
manual.
—
The nosepiece does not move. The nosepiece setting to select
either the 6Position-NP or
5Position-NP on the DIP switch is
wrong.
Specify the correct nosepiece
setting on the DIP switch.
10
Only a specific hole (partial
segment) is used.
Refer to “To extend the life of the
nosepiece” and evenly apply the
grease.
31
Auto-focusing does not work. The safety lock is not attached. Attach the short plug provided
with the product or a lock
mechanism.
—
Repair request
If you cannot improve the phenomena after taking the above remedies, please contact us for assistance.